| The pDR-1500
was developed to meet a need for a fully integrated,
active sampling personal scale instrument with greater
accuracy, increased capabilities, low size and weight,
maximum easeofuse and increased operating time. It was
designed for applications such as site remediation, size
discrimination, mass validation, exposure modeling, and
protection of asthma patients. A lot gets in the way of
accurately measuring aerosol concentration in real-time
– temperature, humidity, air pressure and sample
representation. The pDR-1500 handles all four – with
relative humidity compensation, true volumetric flow
control and legacy pDR nephelometry. An integrated
sample filter enables post-gravimetric validation of
data. Superior particle-cut points compared to those
achievable using impactors are delivered through
volumetric flow control and ACGIH traceable cyclones –
available in pairs, for PM10 and PM4 or PM2.5 and PM1. A
toroidal entrance assures optimized aerosol asporation
and a representative sample even without a cyclone.
|
|
KEY FEATURES AND BENEFITS
- Personal aerosol instrument with
benchtop performance
- Full compensation for
environmental variables
- Interchangeable cyclones for
higher accuracy cut points
- Flexible data logging routines
- Suitable for NIOSH Methods 0500
and 0600
- True volumetric flow control
|
 |
 |
The personalDataRAM™
pDR-1000AN passively measures mass concentrations of
dust, smoke, mists, and fumes in real time, and
sounds an audible alarm whenever the user-defined
level is exceeded.
- Highest performance of any real-time personal
particulate monitor with a measurement range from
0.001 to 400 mg/m3 (auto-ranging).
-
Rugged and
extremely compact, the palm sized pDR-1000AN weighs
only 18 oz. (0.5 kilogram) and can be attached to a
belt or shoulder strap, hand held, operated on a
table top, or mounted on a tripod.
-
Simple zeroing
and calibration – The pDR-1000AN comes
gravimetrically calibrated (NIST traceable) in mg/m3
using standard SAE Fine (ISO Fine) test dust.
Zeroing with particle-free air is accomplished
quickly and effectively under field conditions using
the zeroing kit including with the instrument.
-
Safety approvals and certifications – The pDR-1000AN
complies with US FCC rules (Part 15) and has
received CE certification. It also has intrinsic
safety approval from the US Mine Safety and Health
Administration (MSHA) for use in coal mining
environments containing methane gas. The MSHA Type
2G approval closely resembles the standard intrinsic
safety rating as defined by Class 1, Division 1,
Group D. In addition, it meets US FCC and European
CE rules.
|
|
 
|
|